Contact

Su Zhou Skytest Equipment Co.,Ltd.

Add:No. 122, Xinyangxi Road, Kunshan,Jiangsu, China

Tel:0086-512-50353429 5035 8506

Fax:0086-512-5035 8506

E-mail:sales@szskytest.com

Home > Homemade Products > Low temperature test chamber(SKY6013-225-50)
Low temperature test chamber(SKY6013-225-50)

The low temperature test box complies with gb5013-2008, gb5023-2008 and gbt2951-2008 standard requirements, with the low temperature impact of supporting wire and cable, low temperature winding, low temperature drawing and other devices
Key technical indicators
1. Studio size: 225L (0.5 * 0.6 * 0.75 m);
2. The temperature ranges: - 50 ℃ ~ at room temperature;
3. The temperature uniformity: + / - 1 ℃ or less;
4. The temperature fluctuation degree, plus or minus 0.5 ℃ or less;
Refrigeration system
1. Refrigerating compressor unit: French taikang system;
2. Cooling mode: air cooling type;
3. Environmental refrigerants;
Air conditioning system
1. Air conditioning method: forced circulation ventilation and balance adjustment method. The method not only refers to the refrigeration system under the condition of continuous work, of the temperature control system according to the set point automatically by the PID operation output result to control the output of the heater, eventually reach a dynamic balance;
2. Air circulation device: built-in air-conditioner, circulating air duct and long axial flow ventilator. By improving the air flow, the air flow, heating and cooling capacity are improved, and the temperature uniformity of the test box is greatly improved.
3. Air cooling mode: multistage diaphragm type air heat exchanger;
Enclosure structure
1. Outer wall material: high quality carbon steel plate (1.5mm), surface electrostatic spraying powder treatment;
2. Inner wall material of the box: SUS304 imported a first-class brushed stainless steel plate (1mm);
3. The internal welding of the box has high strength reinforcement;
4. The test case compressor is in split form: it is installed in the rear of the test box and connected with the box through the refrigerating pipe;
5. Insulation: can be used to produce hard polyurethane foam;
Measurement and control system
1. Temperature measurement: high precision (A) thermocouple;
2. Control device: the main controller of the test box adopts the American honeywell IPC2000 high precision double-loop programmable temperature controller;
3. The instrument has the PID parameter self-tuning function;
4. Indicate accuracy: 0.1 ℃;
5. Equipment has independent working time accumulative timer;
Safety protection measures
1. Safe and reliable grounding protection device;
2. Overpressure of refrigerator;
3. Refrigerator overload;
4. Fan overload;
Equipment conditions
1. Power source: 380V plus or minus 10% V, 50Hz, three-phase five-wire system;
2. The environment temperature: 5 ~ 30 ℃;
3. Ambient humidity: less than 85% R.H;
A satisfactory test method
The equipment meets gb2423.1-89, gb2423.2-89 test B: high temperature test method, GB/t2423.3-93 test Ca: constant humidity and heat test method, gb10592-89 high cryogenic tank technical conditions and gb10586-93 humidity and heat test box technical conditions, etc.
[ Print this article ]  [ Back to higher ]  [ Back to top ]

Copyrights Su Zhou Skytest Equipment Co.,Ltd. support: kslcxx.com

Add: No. 122, Xinyangxi Road, Kunshan,Jiangsu, China Tel: 0086-512-50353429 5035 8506 Fax: 0086-512-5035 8506 E-mail: sales@szskytest.com

Online

点击这里给我发消息

点击这里给我发消息