Ozone aging test chamber(SKY6005-100)
Ozone aging test chamber meet GB5013-2008, and GBT2951 GB5023-2008-2008 standard requirement
Technical parameters:
1. The display part: import TFT LCD;
Part 2. Setting: import touch screen control configuration;
3. The workshop size: 400 * 500 * 500 mm (optional) other specifications;
4. The studio is adjustable temperature: RT + 10 ~ 60 ℃;
5. The temperature uniformity, plus or minus 2 ℃;
6. Temperature volatility: plus or minus 0.5 ℃;
7. The temperature deviation: plus or minus 2 ℃;
More than 8. Relative humidity: 60% RH;
9. The humidity control accuracy: 0.1% RH;
10. The ozone concentration: 0 ~ 1000 PPHM direct reading, the deviation of 8% or less;
11. Rack turning speed: 2 r/min (adjustable);
12. Specimen tensile speed: maximum 20 times/minute;
13. The inner material: senior SUS304 stainless steel plate;
14. Carton material: senior SUS304 stainless steel plate;
15. Insulation materials: superfine glass wool;
16. The heater: nickel chromium alloy electric heating wire;
17. The external type small boiler humidification;
18. Special motor, centrifugal fan convection;
19. The water tank water supply, water shortage automatic alarm;
20. Silent ozone generator;
21. The prepared with classic chemical testing interface of ozone concentration, convenient analysis test;
22. The temperature and heat insulation layer adopts hard polyester foam, thickness of 100 mm;
23. The working power supply: AC380V / 50 hz / 20 a.